Product : AGSS System

ANESTHETIC GAS SCAVENGING SYSTM (AGS SYSTEM)
Item Code:-HSS3056
The AGS system shall be a dedicated, specifically designed active extraction and
disposal system for waste anaesthetic gas. It shall provide a maximum flow rate
of 130 l/min with a 1 kPa resistance to flow, and a minimum of 80 l/min with a
4 kPa resistance to flow at each terminal unit, irrespective of the number of
terminal units in use. The AGS system shall use dedicated radial blowers in a
duplex configuration. The AGS pump assemblies shall be skid mounted and
included on the skid shall be the simplex or duplex pump(s), motor control
unit(s) with starter/isolator, moisture drain flask and flexible connector(s) to
connect the plant to the pipeline. Each pump shall include an electric motor and
directly coupled impeller assembly. Impeller bearings in the pump(s) shall not
require lubrication. The pump(s) shall be air cooled and rated for continuous operation.
 